5
General Maintenance
4 Remove the screw securing the filament block to the source mount. (See
Source mount
Screw
Figure 42.
5 Use the hand grip on the filament block to remove the filament block from
Use extra caution when removing the dual filament, as it is extremely brittle.
C AU T I O N
6 Remove the dual filament from the filament block by lifting the source body
136
Figure
42.)
Filament block
Replacing the dual filament
the source.
up off of the filament block, while holding the filament block so that the dual
filament will not fall and become damaged.
Dual filament