Intuvo 9000 GC Column Maintenance
4 Gently check the alignment of the analyzer and the interface.
Forcing the analyzer closed if these parts are misaligned will damage the seal
C AU T I O N
or the interface or the ion source, or keep the sideplate from sealing.
5 You can align the analyzer and interface by wiggling the side plate on its
5975/5977 Series MSD Operating Manual
When the analyzer is aligned correctly, the analyzer can be closed all the
way with no resistance except the spring tension from the interface tip seal.
hinge. If the analyzer still will not close, contact your Agilent Technologies
service representative.
71