Agilent Technologies 5975 Series Operating Manual page 110

Msd for openlab cds
Hide thumbs Also See for 5975 Series:
Table of Contents

Advertisement

4
Operating in EI Mode
Do not overtighten the side plate thumbscrews. Overtightening will strip the
C AU T I O N
threads in the analyzer chamber. It will also warp the side plate and cause
leaks.
8 Plug the MSD power cord in.
9 Switch the MSD on to establish a rough vacuum. Verify that the turbo pump
10 Switch the MSD off.
11 Close the analyzer cover.
12 Disconnect the LAN, remote, and power cables.
The MSD can now be stored or moved. The foreline pump cannot be
disconnected; it must be moved with the MSD. Make sure the MSD remains
upright and is never tipped on its side or inverted.
The MSD must remain upright at all times. If you need to ship your MSD to
C AU T I O N
another location, contact your Agilent Technologies service representative for
advice about packing and shipping.
110
speed is greater than 50%, or that the foreline pressure is 1 Torr.
5975/5977 Series MSD Operating Manual