Agilent Technologies 5975 Series Operating Manual page 66

Msd for openlab cds
Hide thumbs Also See for 5975 Series:
Table of Contents

Advertisement

3
Intuvo 9000 GC Column Maintenance
Ferrule release hole
Transfer line
Bus click and run connection
Figure 16.
12 Remove the 9000 GC/MS Tail from the transfer line and bus. 
Exit in ion source
Figure 17.
13 Replace the gasket. (See
14 Carefully slide the 9000 GC/MS Tail into the GC/MS interface, and gently
15 Check that the 9000 GC/MS Tail click and run fitting sits flat against the
16 Screw the 9000 GC/MS Tail column nut to the transfer line threaded
17 Loosely install the two compression bolts.
66
MS tail heater assembly clamp open with 9000 GC/MS Tail removed
If the ferrule becomes stuck, press a pointed object, like a paper clip, into the
ferrule release hole at the end of the transfer line.
9000 GC/MS Tail removed from GC/MS Interface
page 59.)
place the click and run connection into the bus connection.
gasket in the bus connection.
connector finger tight, then use a ¼-inch wrench to tighten the nut an
additional 20 to 30 degrees.
Ferrule and nut
Click and run connection
"To Replace an Intuvo 9000 GC Gasket"
5975/5977 Series MSD Operating Manual