Technical Data

Type / Maximum capacity (E
Maximum capacity (E
Accuracy class
Maximum number of load cell
intervals (n
)
LC
Min. LC verification interval (v
Sensitivity (C
)
n
Temperature effect
on sensitivity (TK
)
C
Temperature effect
on zero balance (TK
Hysteresis error (d
hy
Non-linearity (d
)
lin
Creep (d
) over 30 min.
cr
Input resistance (R
LC
Output resistance (R
Reference excitation voltage (U
Nominal range
of excitation voltage (B
Insulation resistance (R
Nominal temperature range (B
Service temperature range (B
Storage temperature range (B
1)
The data for Non--linearity (d
typical values. The sum of these data meets the requirements according to OIML R60.
A1153--1.2 de/en/fr
)
max
)
max
)
min
of E
% of C
% of C
)
0
)
% of C
)
)
0
)
ref
)
U
)
is
)
T
)
tu
)
tl
), Hysteresis error (d
lin
RTN 0.05
0.05
--
%
--
max
mV/V
0.05
n
n
/ 10K
0.03
0.05
0.05
n
n
0.03
4010 2
V
GΩ
--10 ... +40 [+14 ... +104]
--40 ... +80 [--40 ... +176]
°C [°F]
(Option: up to +110 °C / +230 °F)
--50 ... +85 [--58 ... +185]
) and Temperature effect on sensitivity (TK
hy
RTN C3
1 t ... 470 t
C3 (OIML R60)
3000
0.005
(= E
/ 20000)
max
2.85 0.1 %
0.008
0.007
0.02
0.02
0.017
4450 100
4010 0.5
5
5 ... 30 (max. 60)
> 20
25
1)
1)
1)
) are
C
HBM