EMS AIR-FLOW MASTER Perio Operation Instructions Manual page 38

Device for dental prophylaxis
Table of Contents

Advertisement

Use only powder provided by EMS and corresponding to the application.
Close correctly the bottle after use. The conditions of storage are described in the chapters "Storing
the product" and "Technical data".
A set of interchangeable colored rings is provided to help recognize the chosen powder flavors. Please
refer to the information provided in the chapter "Maintenance".
Check the condition of the threads on the powder chamber, on the cap and on the O-ring seals.
Please refer to the information provided in the chapter "Maintenance".
Clean the threads of the powder chamber and of the cap before screwing them on.
Ensure that the powder chambers are closed tight. Otherwise air and powder could leak.
Do not turn the powder chambers upside down after filling.
The PERIO powder being specially fine, tamping it down the powder chamber must be avoided
when filling it.
The powder should not remain over a determined time inside the powder chamber. Read the
instructions delivered with the powders.
The PERIO powder level will go down slightly a few minutes after the filling (powder compaction).
26
FB-580_CN_rev_B_01_Not Util AIR-FLOW MASTER.indd 26
13.10.2017 09:59:20