Parameter Settings

5-8 | 303B Moisture Analyzer Manual
In step 12, P x.xx represents your software revision number, which can be
helpful to identify software upgrades. This should be the same number
written on the cpu/display board cpu PROM.
Offset
Offset subtracts from the moisture readings.
1.
Press the POWER key (LED OFF). Hold the STANDBY key (LED
ON) and press the POWER key (LED ON) until the STANDBY, +, and –
LED's flash.
2.
Press the + or – key to increment or decrement the display to the
desired valve.
3.
When the desired value is displayed, press OFFSET to store the
value; ProG will be displayed until the value is stored, followed by the
normal LED/display test.
If the ALARM, CELL TEST, 0-1000 (50), 0-2000 (100), or DISPLAY
NOTE
LIGHT key is pressed during the routine, the new value is not stored
and normal moisture measuring begins.

Gain

Gain adjustment enables a value to be entered to compensate for non-
standard ranges or flow limitations.
1.
Press the POWER key (LED OFF). Hold the ELEC TEST key (LED ON)
and press the POWER key (LED ON) until the STANDBY, +, and –
LED's flash.
2.
Press the + or – key to increment or decrement the display to the de-
sired valve.
3.
When the desired value is displayed, press GAIN to store the value;
ProG will be displayed until the value is stored, followed by the nor-
mal LED/display test.