Factory A/D Gain - Ametek 303B User Manual

Moisture monitor
Table of Contents

Advertisement

NOTE
If the ALARM, CELL TEST, 0-1000 (50), 0-2000 (100), or DISPLAY
LIGHT key is pressed during the routine, the new value is not stored
and normal moisture measuring begins.

Factory A/D Gain

Factory A/D gain linearizes the cpu/display board to the moisture cell.
1.
Press the POWER key (LED OFF). Hold both ELEC TEST and STAND-
BY keys (LED's ON) and press the POWER key (LED ON) until the
STANDBY, ELEC TEST, + and - LEDs flash.
2.
Press the + or - key to increment or decrement the display to the de-
sired value.
3.
When the desired value is displayed, press STANDBY to store the
value; ProG will be displayed until the value is stored, followed by the
NOTE
If the ALARM, CELL TEST, 0-1000 (50), 0-2000 (100), or DISPLAY
LIGHT key is pressed during the routine, the new value is not stored
and normal moisture measuring begins.
|
Maintenance
5-9