Page 2-10
Calibrating the Barometer
The adjustment range for the barometer is the displayed value ± 20 mmHg.

Recalling Last Sample Data

The 248 retains the data for the last sample measured. The Data Recall
menu allows:
• data to be recalled,
• patient data to be entered or changed,
• results to be printed.
1. From Ready press # for menu, 4 for Data Recall and 1 for Last Sample
2. Press 1 to recall the results Screen 1 to the display. (The calculated
3. Press 2 if you want to enter or change the patient data. You can enter
4. Press 3 if you want to print the data. The printout shows the date and
→ Calibration → Barometer
Enter atmospheric pressure
press * to Exit
Data.
→ Data Recall → Last Sample Data
1 View Results
2 Enter Patient Data
3 Print Data
press 1 - 3 or * to Exit
parameters are also recalled, press # to see Screen 2.)
patient and operator ID, and patient temperature, ctHb and F
You can enter up to 12 digits for operator and patient ID. You can use
the decimal point to insert dashes.
The other values that can be entered are:
• patient temperature 10.0 - 43.9°C
• ctHb 2.0 - 25.0 g/dL (20 - 250 g/L or 1.2 - 15.5 mmol/L)
• F
O
15.0 - 100.0%.
I
2
The 248 will use these values for recalculating the results. If you do not enter
patient data the 248 uses the values entered previously, or if none were
entered normal (default) temperature (37°C), ctHb (15 g/dL (150 g/L, 9.6
mmol/L)) and F
O
(20.9%) are used, with the following exceptions:
I
2
• O
CT will not be reported unless ctHb is entered, and
2
• pO
(A-a) and pO
(a/A) will not be reported unless F
2
2
time the sample was analyzed, not the time of recall.
760
mmHg
248 Operator's Manual
O
.
I
2
O
is entered.
I
2