Waste Disposal - Siemens RAPIDLAB 248 Operator's Manual

Table of Contents

Advertisement

Understanding the System
Active Ingredients
Material Safety Data Sheets for the 248 reagents are supplied by your
local distributor.
Intended Use
7.382 buffer
6.838 buffer
Wash
Deproteinizer removes protein buildup from the sample path. It is used
Conditioner
Storage
Store all reagents away from direct sunlight at 4 - 25°C.
Discard 7.382 and 6.838 buffers 30 days after opening.
Do not use reagents after the expiry date.
Discard Deproteinizer and Conditioner solution after single use.
Handling and Preparation
7.382 buffer, 6.838 buffer, Wash and Conditioner require no
preparation before use.
Prepare Deproteinizer as directed by the instructions on the package.

Waste Disposal

BIOHAZARD:
Refer to Page 3-4, Emptying the Waste Bottle, for detailed instructions
on handling the waste bottle.
Discard the waste bottle and its contents according to your laboratory
protocol. NCCLS Publication GP5 gives detailed guidelines.
provides the calibration point for 1 and 2 point pH
calibrations. 7.382 buffer is buffered to a pH of 7.382 at
37°C and is NIST traceable.
provides the slope point for 2 point pH calibrations. 6.838
buffer is buffered to a pH of 6.838 at 37°C and is NIST
traceable.
washes the probe and sample path.
regularly as preventative maintenance for the 248.
cleans and conditions the pH sensor. It is used regularly
as preventative maintenance for the 248.
See Appendix A, Protecting Yourself From Biohazards.
Page 1-11
25