Operating the System
Analyzing Syringe Samples
6. Press # if you want to enter patient data.
The 248 will use the values entered for patient temperature, ctHb and F
when calculating the results. If you do not enter patient data the 248 uses
normal (default) temperature (37°C), ctHb (15 g/dL (150 g/L, 9.6 mmol/L))
and F
Results may also be corrected for patient temperature, ctHb and F
measurement - see Recalling Last Sample Data, Page 2-10.
Example result. If the measured values are outside the reference ranges
an arrow indicates if they are above or below the range. If you selected
calculated parameters they are displayed on Screen 2. (See Section 5,
Configuring the System, for details on reference ranges and calculated
parameters.)
If a bubble or short sample is detected the 248 will alert you.
See Page 2-8 for detalis.
Measuring
pH
pCO
mmHg =
2
pO
mmHg =
2
#
press
to enter patient data
Enter Patient Data
Operator ID - - - - - - - - - - - - Temp - - . -
Patient ID
- - - - - - - - - - - - ctHb
#
press
for next entry or * to Exit
You can enter up to 12 digits for operator and patient ID. You can use
the decimal point key to insert dashes.
The other values that can be entered are:
• patient temperature 10.0 - 43.9°C
• ctHb 2.0 - 25.0 g/dL (20 - 250 g/L or 1.2 - 15.5 mmol/L)
• F
O
15.0 - 100.0%.
I
2
O
(20.9%) values in the calculations, with the following exceptions:
I
2
• O
CT will not be reported unless ctHb is entered, and
2
• pO
(A-a) and pO
2
2
Results - Screen 1
pH
pCO
mmHg =
2
pO
mmHg =
2
#
press
to see Screen 2 or * to Exit
=
7.392
41.7
78.4
FIO
(a/A) will not be reported unless F
Measured
at 37°C
=
7.392
41.7
78.4 ↓
°C
- - . -
g/dL
- - - . -
%
2
O
is entered.
I
2
O
I
2
Corrected
to 33.5°C
7.444
35.8
62.2 ↓
Page 2-3
O
I
2
after