Serial Data Transmission Of Temperature Data - Siemens MPA 1 Series Instruction Manual

Spectral pyrometer ardometer
Table of Contents

Advertisement

s
Most key parameters are directly available in the main menu. Further
functions are contained in submenus. Navigation within the menus is ex-
plained in the following graphics:
Enter
Select [KEY]
Enter
[ESC]
Select [KEY]
Enter
[ESC]
To set the pyrometer to the terminal mode, simultaneously hold down
the Ctrl key and press the E key twice in rapid succession.
Direct commands have an assigned key. Example: E for emissivity (epsi-
lon). Submenu settings are shown in brackets. Example: [LAMBDA 1]
13.1

Serial Data Transmission of Temperature Data

Transmission parameters for the serial interface:
57600 Baud / 8 data bits / odd parity / 1 stop bit / no handshake
Temperature format (one cycle):
Byte
Negative Temperature
1
Minus symbol "-"
2
Digit 1000
3
Digit 100
4
Digit 10
5
Digit 1
6
Decimal point "."
7
Decimal place
8
Space
9
Unit "C" or "F"
10
Space
11
Carriage Return
Main Menu
[RETURN]
or [ESC]
Backspace
Submenu 1
Backspace
Submenu 2
Positive Temperature
Space
Digit 1000
Digit 100
Digit 10
Digit 1
Decimal point "."
Decimal place
Space
Unit "C" or "F"
Space
Carriage Return
40
Operating Manual MPA
press
Execute the
command
press
Execute the
[RETURN]
command
or [ESC]
press
Execute the
[RETURN]
command
or [ESC]
Temperature falls below
measuring range
Space
Minus symbol "-"
"U"
"N"
"D"
"E"
"R"
Space
Minus symbol "-"
Space
Carriage Return
Temperature exceeds
measuring range
Space
Minus symbol "-"
"O"
"V"
"E"
"R"
Space
Space
Minus symbol "-"
Space
Carriage Return