Replacing The Printer Paper - Siemens RAPIDLAB 248 Operator's Manual

Table of Contents

Advertisement

Maintaining the System

Replacing the Printer Paper

Equipment: Printer paper (Cat. 673252).
CAUTION:
print quality or damage the printer.
Replace the printer paper when the red stripe appears or when
prompted by the display.
1. Tilt the paper cover backwards.
2. Tear off any remaining paper, and remove the old paper roll.
3. Make sure the end of the new paper roll is square. Hold the new
CAUTION:
in incorrectly the printer will not print, and it may cause paper jams.
Figure 3-26. Loading the Printer Paper
4. Feed the paper down into the paper slot, and press the
Figure 3-27. Feeding the Printer Paper into the 248
5. Place the paper roll in the holder. Close the paper cover.
6. Press the
7. Test the printer, to check that the printout is clear, Page 4-13.
Only use Siemens printer paper. Other paper may affect
paper roll in one hand with the paper coming from the bottom of the
roll and towards you. Bend the end of the paper back slightly.
Make sure you feed the paper in correctly. If the paper is fed
until the feed mechanism takes up the paper and pulls it.
key again until the paper appears through the tearer
at the front of the 248.