Maintaining the System
when maintenance has been carried out.
Daily
• check calibrant levels and 'change by' date, shake buffer pack • wipe sample area, drip tray and external
surfaces • check drip tray is fitted and connected properly •
Month
1
2
3
4
Weekly
• check fill solution in pH and reference sensors, and debubble sensors • deproteinize/condition sensors
• check gas cylinder pressure and flow rate •
1
14
15
27
28
40
Quarterly
• Use Disinfect routine • replace pump tubing, clean and lubricate rollers • replace drip tray if difficult to
clean •
Wk 13
Six-monthly
• replace bottle tubing •
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20
2
3
4
16
17
18
29
30
31
41
42
43
44
Wk 26
5
6
7
19
20
21
32
33
34
45
46
47
Wk 26
Start Date . . . . . . . . . . . . . . .
21 22 23 24 25 26 27 28 29 30 31
8
9
10
11
22
23
24
35
36
37
48
49
50
Wk 39
Wk 52
Page 3-3
12
13
25
26
38
39
51
52
Wk 52