Installation Instructions; Suggested Tools; Making Gas Connections - Ametek ta3000R User Manual

Gas purity monitor
Table of Contents

Advertisement

24 | Model ta3000R Gas Purity Monitor

INSTALLATION INSTRUCTIONS

The installation instructions must be strictly followed for safe and
!
reliable operation of the analyzer.
CAUTION
If any inconsistencies are encountered during installation, stop im-
!
mediately. Do not continue the installation without direction from
Ametek customer service.
CAUTION
Tools Suggested for Installation
All connections are made using U.S. standard size conectors.
Open end wrenches: 1/4", 3/8", Phillips head screwdriver
Flow Meter (Digital in the sccm range)
Mirror
Magnet

Making Gas Connections

Prepare for Gas Connections
Remove the red plastic shipping cap from the Hg vapor trap (scrubber).
Remove any 1/16" VICI plug from the rear or front panel(s). Remove all
other shipping caps immediately prior to connection of gas supplies.
Connect Carrier Gas (Actuator Gas Supply)
The ta3000 analyzer uses a chromatographic carrier gas that is also used
internally as actuator gas. Matching the carrier gas to the sample will
prevent upsets caused by the elution of the extraneous balance peak. See
gas specifications for Carrier Gas on page 14, Specifications for details. The
instrument connection is made at the "CARRIER" 1/16" VICI port on rear
panel. Pressurize external line to 80 psig (Minimum 70 psig, Maximum 90
psig).
Connect Sample Gas without Front Injection Panel option
The ta3000 analyzer can be used for process or single gas sample analysis.
On-line process sample line should be connected to the "SAMPLE IN"
1/16 " VICI fitting on the rear panel. Optimal results are achieved with
constant flow rates (20-100 sccm).