Diagnostic Print Key

The Print key appears when setting the clock. Pressing the Print key from
this page will print a complete report of all system settings, calibration
settings, temperatures and status and the chromatogram from the last
run of the analyzer. This is a system diagnostic tool which can be used to
troubleshoot problems with the analyzer.
Saving changes into Non-Volatile RAM
After changes have been made to any of these fields, the system LED will
turn red. This indicates that there is a system error or system state change.
Use the Page key to move to the System Settings page. Any change made
will be shown under the System Errors line. Use the arrow keys to move
the cursor in front of the change you wish to Accept and press the key un-
der the Accept menu label. Next, press the Save key to save the accepted
changes into Non-Volatile RAM.
The instrument must be in "Idle" mode to accept the changes.
!
CAUTION
Temperatures and Status Page
Press the Page key to display the Temperatures and Status page. This page
shows status information about the critical heated zones and status of the
detector. No user input is allowed on this page in normal user mode.
Image of Temperatures and Status Screen
Any change made will be lost if the instrument is turned off before the
changes have been accepted on the System Settings page.
Idle Mode
Temperatures & Status
RGD Bed =
RGD Column =
RGD Lamp = On
RGD Vout = 36187
RGD Vsig = 1978.4
------------------------- Pressure --------------------------
Carrier = 20.3
12:34:25
Meas
Set
265 #
265
128 #
130
Page
User Manual | 33