Calibration - Ametek ta3000R User Manual

Gas purity monitor
Table of Contents

Advertisement

CalibrationScreen
Calibration Screen
34 | Model ta3000R Gas Purity Monitor
Press the Page key to display the calibration information page.
The calibration page is used to calibrate the instrument and shows
information about the most recent run. Name is the gas species to be
measured. Conc. is the concentration (or blend) of the calibration gas
connected to the instrument. Area is the peak area of the impurity
detected during the last analysis run for each impurity type.
Rfactor is the response factor calculated by the system on the most
recent calibration. Rfactor is used to calculate the concentration of
impurities. RT is retention time of the impurity. This is the time corre-
sponding to the top of each peak calculated from the start of the most
recent analysis run.
Idle Mode
Name
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
H2
*
CO
Pressing page again will exit the page screens.
It is recommended to use the Ametek's/Trace Analytical MGB1000
Micro Gas Blender to blend the calibration gas to the desired level
when calibrating the ta3000R.

Calibration

Conc.
Area
Rfactor
50
15025
50
62341
Conc. Factor 1
Edit
12:34:25
RT
Page