This chart illustrates how the menus are arranged in Super User Mode.
Each large circle represents a screen that is displayed on the instrument.
The names printed above each arrow represent a menu label and menu
key. The arrows indicate where pressing the key associated with that label
will take the user.
Shut Down
System shut down for maintenance or for any other reason is very simple.
If shut down is due to an emergency or for any safety reason, simply
switch the instrument off using the A/C DIN Switch located on the rear
panel of the analyzer (0 depressed is off).
!
CAUTION
To shut down the analyzer:
From the main menu press the Idle key to stop analysis.
Shut off any support gasses except for the carrier gas.
Press the Cool key and allow the analyzer to cool down for two to three
hours.
Switch the analyzer and all other components off.
Shut off the carrier gas.
Carrier gas to the instrument must not be stopped until all tempera-
0
tures are below 50
C.
User Manual | 39