Guidance and Manufacturer's Declaration – Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below.
The customer or the user of this device should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transients/bursts
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE 1: U
is the a.c. mains voltage prior to application of the test level.
T
NOTE 2: r.m.s.: root mean square
*1: Not applicable because it does not connect directly to outdoor cable.
38
IEC 60601 Test Level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
±1 kV for input/output lines
AC/DC power
±0.5 kV, ±1 kV line(s) to line(s)
±0.5 kV, ±1 kV, ±2 kV line(s) to
earth
Signal input/output
±2 kV line(s) to earth
dips
0% U
: 0.5 cycle (at 0, 45, 90,
T
135, 180, 225, 270, 315°)
0% U
: 1 cycle (at 0°)
T
70% U
: 25/30 cycles (at 0°)
T
25 (50 Hz)/30 (60 Hz)
short interruptions
0% U
: 250/300 cycles
T
250 (50 Hz)/300 (60 Hz)
30 A/m (r.m.s.)
50 Hz or 60 Hz
Compliance Level
±2 kV, ±4 kV, ±6 kV, ±8 kV
contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
±1 kV for input/output lines
AC/DC power
±0.5 kV, ±1 kV line(s) to line(s)
±0.5 kV, ±1 kV, ±2 kV line(s) to
earth
Signal input/output
*1
±2 kV line(s) to earth
dips
0% U
: 0.5 cycle (at 0, 45, 90,
T
135, 180, 225, 270, 315°)
0% U
: 1 cycle (at 0°)
T
70% U
: 25/30 cycles (at 0°)
T
25 (50 Hz)/30 (60 Hz)
short interruptions
0% U
: 250/300 cycles
T
250 (50 Hz)/300 (60 Hz)
30 A/m (r.m.s.)
50 Hz or 60 Hz
Electromagnetic Environment
– Guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
If user of this device requires
continued operation during
power mains interruptions, it is
recommended that this device be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic field
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Operation 2018-10-22