Positioning the patient
1
Have the patient wear an X-ray protection apron.
Set radiography media (film, sensor, and imaging plate) in the patient's mouth correctly.
Have the patient seated when R-Type is used.
Store radiography media in a clean, uncontaminated area.
Poor patient positioning may make it hard to make a correct diagnosis using the captured image.
Pole Brace
• Infection control procedures must be observed when using accessories, such as film holders, X-ray tube guides, imaging plates and CCD detectors. When us-
ing accessories always follow the manufacturer's instructions on how to use the accessory and prevent cross contamination from one patient to another.
• When lowering the pole, do not pinch your fingers between the pole and the main support column.
• Maximum loading limit of Body Weight is 135 kg. Do not use this equipment for patients who weight more than 135 kg.
• Have patients remove glasses, earrings and any other accessories that might interfere with making a good exposure.
• When moving the light pole up or down or rotating the arm, do not use excessive fore, especially if the pole or arm reaches its stopper. This could damage the
equipment. (R-Type)
12
Pull up the seat adjustment
lever to adjust the seat height.
Slide the pole up or down to adjust the height of the X-ray head.
Operation 2019-04-22