Morita Veraview iX Operation Instructions Manual page 23

Dental x-ray unit
Table of Contents

Advertisement

Timer table
No. 10
Adult (M)
Lady (S)
Child (C)
No. 11
Adult (M)
Lady (S)
Child (C)
No. 12
Adult (M)
Lady (S)
Child (C)
No. 13
Adult (M)
Lady (S)
Child (C)
No. 14
Adult (M)
Lady (S)
Child (C)
Recommended values (*For 60 kV)
Radiography condition
Film (Sensitivity D)
Film (Sensitivity E)
Film (Sensitivity F)
Digital (IP)
Operation 2017-02-24
Upper
Upper
incisor
premolar
0.40
0.50
0.25
0.32
0.20
0.25
0.50
0.63
0.32
0.40
0.25
0.32
0.63
0.80
0.40
0.50
0.32
0.40
0.80
1.00
0.50
0.63
0.40
0.50
1.00
1.25
0.63
0.80
0.50
0.63
Standard cone
No. 11 (F. 11)
No. 08 (F. 08)
No. 05 (F. 05)
No. 07 (d. 07)
Upper
Lower
molar
incisor
premolar
0.63
0.25
0.40
0.16
0.32
0.13
0.80
0.32
0.50
0.20
0.40
0.16
1.00
0.40
0.63
0.25
0.50
0.20
1.25
0.50
0.80
0.32
0.63
0.25
1.60
0.63
1.00
0.40
0.80
0.32
Timer table (display)
Lower
Lower
Occlusion
molar
0.32
0.50
0.25
0.32
0.16
0.25
0.40
0.63
0.32
0.40
0.20
0.32
0.50
0.80
0.40
0.50
0.25
0.40
0.63
1.00
0.50
0.63
0.32
0.50
0.80
1.25
0.63
0.80
0.40
0.63
(Unit: sec.)
Long cone
No. 14 (F. 14)
No. 11 (F. 11)
No. 08 (F. 08)
No. 10 (d. 10)
0.80
0.50
0.40
1.00
0.63
0.50
1.25
0.80
0.63
1.60
1.00
0.80
2.00
1.25
1.00
21