HIGH MESSAGE:
MEANING: Glucose test result is above 600 mg/dL. This high
result may indicate hyperglycemia (high blood glucose). The
high result is stored in the meter. On the app it will be saved
as a test result of > 600 mg/dL with time and date. The value
will be represented as 600 mg/dL in the statistics in the app.
ACTIONS: If you feel symptoms such as fatigue, thirst,
excessive urination, or blurry vision, follow your healthcare
professional's recommendations. If you get a high glucose
test result but have no symptoms of high blood glucose,
then retest with a new BGStar Test Strip. If you still get a high
glucose test result, follow your healthcare professional's
recommendations. Checking ketones may be advisable.
LOW TEMPERATURE MESSAGE:
MEANING: The iBGStar Blood Glucose Monitoring System
is below its operating temperature range of 50°F to 104°F
(10°C to 40°C).
ACTIONS: Move to an area with an ambient temperature of
50°F to 104°F (10°C to 40°C). Wait for the meter and test strips
to reach the new temperature (usually 10-20 minutes) before
using the meter or performing a test.
HIGH TEMPERATURE MESSAGE:
MEANING: The iBGStar Blood Glucose Monitoring System
is above its operating temperature range of 50°F to 104°F
(10°C to 40°C).
68
ACTIONS: Move to an area with an ambient temperature of
50°F to 104°F (10°C to 40°C). Wait for the meter and test strips
to reach the new temperature (usually 10-20 minutes) before
using the meter or performing a test.
LOW BATTERY MESSAGE:
This message appears for 2 seconds.
MEANING: The battery in your meter is low on power.
ACTIONS: If the battery symbol appears, the battery in your
meter must be charged. For details about charging your
meter, refer to Chapter 17.
DEAD BATTERY MESSAGE:
MEANING: The battery in your meter does not have enough
power to activate your meter and perform a test.
ACTIONS: If the dead battery symbol appears, the battery in
your meter must be charged. For details about charging your
meter, refer to Chapter 17.