[6] Viewing Control Solution Test Result:
The control solution test result will appear on the meter display.
The unit of measure (mg/dL), time and date of testing will
scroll to the right of the control solution test result. The
control solution symbol will appear next to the test result. The
control solution test result is stored in the meter's memory.
Example Results:
In addition, if the meter is connected to an iPhone or iPod
touch and the app is launched, the control solution test result
scorecard will appear on the iPhone or iPod touch's screen.
Compare the result of your control solution test to the range
printed on your BGStar Blood Glucose Test Strip vial label. Be
sure you look at the range for the control solution level (Level
2 or Level 4) that you used. Your control solution result should
fall within this range.
EXAMPLE:
38
EXAMPLE:
Level 2: 90-150 mg/dL
Level 4: 220-320 mg/dL
FOR EXAMPLE ONLY.
Refer to your test
strip vial label for
corresponding range.
If your control solution test results are out of the range, repeat
the test and refer to Chapter 8 of this guide to see potential
causes of error. If the results from the control solution tests
continue to fall outside that range, do not use the iBGStar
Blood Glucose Monitoring System to test your blood glucose.
Call Customer Service at 1-800-633-1610. The phone number
is also printed on your iBGStar meter's cap label.
[7] Remove the Used BGStar Test Strip From the iBGStar
Meter's Test Strip Port:
The control solution test result scorecard will be downloaded
to the app on your iPhone or iPod touch. The battery status
and charging battery symbol will appear on the meter's display.
Example Results:
When using your iBGStar meter independently, removing
the used test strip will deactivate your meter. When using
your meter with the app closed, removing the used test strip
will make your meter display the battery status and charging
symbol.
All test result scorecards in the app that have been
downloaded from the iBGStar meter will have a Lock
icon in the upper left corner (in place of the Delete
icon). The Lock icon indicates that the control solution value
and associated time and date cannot be deleted or edited.
39