Important Information About Performing Control Solution Tests - AgaMatrix iBGStar Owner's Manual

Table of Contents

Advertisement

7
Important Information About Performing
Control Solution Tests
The iBGStar Blood Glucose Monitoring System can be used
with BGStar Control Solution (control solution).
Control solution should only be used for 90 days after first
opening the bottle or until the expiration date printed on the
label, whichever comes first.
IMPORTANT: Count forward 90 days from the date you open
a new bottle of BGStar control solution. This is your discard
date. Write the discard date on the BGStar control solution
bottle.
CAUTION: Results from control solution tests do not refl ect
your blood glucose level. The control solution range is a
target range for control solution only. It is not a target range
for your blood glucose level.
The control solution is used to:
[1] Ensure that your iBGStar Meter and BGStar Test Strips (test
strips) are working together properly.
[2] Practice testing without having to use your own blood.
30
Control solution tests should be performed when you:
[1] First get your meter.
[2] Suspect that your meter or test strips are not working
properly.
[3] Think your test results are not accurate.
[4] Have dropped or damaged your meter or exposed your
meter to liquids.
[5] Are advised by your healthcare professional to do so.
Important BGStar Control Solution information:
[1] Use only BGStar Control Solution with your iBGStar Blood
Glucose Monitoring System.
[2] Always shake the control solution bottle before use.
[3] Always discard the first drop of control solution and wipe
the bottle tip clean using a clean tissue or paper towel.
[4] Replace the cap on the control solution bottle immediately
after use.
[5] Do not add water or any other liquid to control solution.
[6] Control solution tests should be performed within the
system operating temperature range of 50°F to 104°F (10°C
to 40°C).