Download Print this page

Troubleshooting - Emerson Rosemount Wireless Pressure Gauge Quick Start Manual

With wireless hart protocol

Advertisement

Quick Start Guide
6

Troubleshooting

This section provides information for basic troubleshooting. See
Pressure Gauge Reference Manual
Device status
The flashing LED indicates device status using the colors descibed.
Table 6-1: Status Descriptions
LED color
6.1
Pressure measurement
If the mounting effects have not been eliminated after completing
mounting
pressure value.
Procedure
1. From the HOME screen, eter the HART
Device dashboard Fast
Keys
2. Follow the commands to perform the procedure.
6.2
Wireless connectivity
If the device has not joined to the network after power up, verify the
following:
Active Advertising has been enabled on the Smart Wireless Gateway
14
Device status
Green
Functioning properly
Amber Battery is low, battery replacement recommended
Red
Battery replacement required OR Device is
malfunctioning
No
No power, verify ON/OFF switch is in "on" position
color
effects, perform this alternative procedure for verifying the
for advanced troubleshooting.
®
Fast Key sequence.
2, 2, 1, 1, 1
Feburary 2021
Wireless
Eliminate
Emerson.com/Rosemount

Advertisement