Chapter 10
Notes on registering calibration settings
 If the nosepiece is not set at a click-stop position, calibration is disabled.
 If the AZ100M microscope is connected to the DS-L4, the DS-L4 automatically recognizes the AZ-ICI coaxial
illuminator attached to the microscope. If the AZ-ICI is attached, the DS-L4 measurement function
automatically multiplies each measurement value by 1.25x, the magnification of the objective, instead of using
the calibration setting. For the magnification of the zoom/intermediate magnification lens, enter 0.6x, the
magnification of the vertical tube of the AZ100M. Do not enter the magnification of the coaxial illuminator.
 If the SMZ1270i or SMZ18 microscope is connected, depending on the position of the zoom knob, the DS-L4
may not be able to detect the zoom magnification correctly. As a result, the displayed measurement value may
vary constantly. In such a case, rotate the zoom knob until variation stops.
(4)
Setting the magnification of the relay lens
You can set the magnification of the relay lens attached to the TV adapter of the microscope. Note that the magnification
set here is not applied when MANUAL calibration mode is selected.
1
From the [TOOL MENU: MEAS/DRAW] screen, display
[TOOL SET]. Alternatively, from the [TOOL MENU:
SCALE] screen, display [TOOL SET].
2
Tap [RELAY LENS].
3
Tap [SELECT], and then select the magnification of the
relay lens that is being used.
4
Tap [X] to close the [TOOL SET] screen.
To edit the list of relay lenses
(1) Tap [EDIT], and then tap the button for the lens
whose data you want to edit.
(2) Enter the magnification.
(3) Tap [SAVE].
Preparation for Using the Annotation Function and Measurement Function
103
3
2
Setting the magnification of a relay lens