Nikon DS-L4 Instructions Manual page 109

Microscope camera control unit
Hide thumbs Also See for DS-L4:
Table of Contents

Advertisement

Chapter 10
If the selected calibration setting is deleted
If the calibration setting that is currently selected is deleted by using the [MEAS/DRAW] or [SCALE] menu, the
DS-L4 behaves as follows.
 If the microscope is connected to the DS-L4 using a USB cable, the calibration value is initialized with the
information of the objective that has been registered in the microscope ([MIC CALIB]). For details on
registering the information about the objectives with the motorized microscope, refer to the documentation for
the microscope or its setup tool. Also, refer to "10.1.3 (3) Switching the calibration setting".
 If the microscope is connected to the DS-L4 without using a USB cable, calibration is disabled.
(3)
Switching the calibration setting
If the microscope is connected without using a USB cable
1
Display the [TOOL MENU: MEAS/DRAW] or [TOOL
MENU: SCALE] screen, and tap [CALIB]. Alternatively,
tap the desired measurement or drawing button, and
then tap [CALIB].
2
Select the desired calibration setting.
 OFF: No calibration setting is used.
 [M1] to [M14]: Local calibration settings that have
been registered
The icon for the selected calibration setting is displayed.
Switching of the calibration setting and overlay information
Switching of the calibration setting affects the scale,
annotation, and measurement functions as follows:
 The scales (scale bar, cross-hairs, crossed scales, and
bars for XY measurement) are redrawn in accordance with
the new calibration setting.
 The annotations do not change.
 The measurement results are updated in accordance with
the new calibration setting.
Comments on calibration settings
Add a comment for each calibration setting.
In the [CALIBRATION] submenu, the comments
entered when calibration settings are registered are
displayed on the icons for the [M1] to [M14] settings.
When one of the [M1] to [M14] settings is selected, the
comment on the selected setting is displayed in the
command display areas at the bottom of the [CAMERA
MENU] and [INFO: CAMERA] windows.
If a motorized microscope is connected to the DS-L4 with a USB cable
When the nosepiece is rotated, the DS-L4 detects the address of the selected objective to switch the calibration setting
automatically.
Make sure that the information about the objectives has been registered in the motorized microscope. The information
about the objectives is used as calibration settings. (To check the information about the currently selected objective, tap
[CALIB] on the DS-L4.) For details on how to register the information about the objectives, refer to the documentation for
the microscope or its setup tool.
Preparation for Using the Annotation Function and Measurement Function
101
2
1
Switching the calibration setting
Comment
entered at
the time of
registration
Comment on a calibration setting