Operation
Dew formation can also be prevented by using the humidifier delay control option. See the
table below for some typical dew points. For example, when temperature and humidity
inside the chamber are 85°C and 85% RH respectively, the dew point of the air inside the
chamber will be 80.9°C. Therefore, dew will form on specimens if their surface
temperature is less than 80.9°C.
Dry-Bulb
TABLE 6.13.
Relative Humidity
(% RH)
60
85
85
70
90
85
85
90
Platinous User's Manual
D
EW POINT
Dew Point (°C)
56.5
66.3
67.7
80.9
82.3
65