Troubleshooting
Displayed Alarm
Code/Channel
Code (option)
COMMUNICATION
The instrumentation
ERROR-5
detected system trouble
(99)
in the CPU board. The
chamber has been
stopped.
COMMUNICATION
The instrumentation
ERROR-6
detected system trouble
(99)
in the CPU board. The
chamber has been
stopped.
94
Trouble
Refrigeration task
error on CPU
board.
Temperature
control task error
on the CPU
board.
Platinous User's Manual
Cause
Remedial Action
Reactivate the system from
the main power switch
(leakage breaker) and resume
testing. If the same alarm
occurs again, call for service.
Reactivate the system from
the main power switch
(leakage breaker) and resume
testing. If the same alarm
occurs again, call for service.