Quality Assurance
Positioning aid
Dosimeter
syngo CT 2007E
02/2008
To position the 16 cm CTDI head phantom use the CFK head
F . 4
holder and, if necessary, the supplied thin pads to immobilize it.
The 32 cm CTDI body phantom has to be placed on the patient
table and fixed with the pads, if necessary.
Measurement is performed with the dosimeter, a dose measur-
F . 4
ing chamber with the necessary accessories, and a thermome-
ter.
The dosimeter is not supplied with the system.