Scan Unit (Gantry)
Table retraction
STOP pushbuttons
B.2-12
syngo CT 2007E
02/2008
B . 2
With this key you move the table top out of the gantry opening
and lower it. The button is active as long as it is pressed. The
patient can then leave the table.
B . 2
The gantry is automatically set vertical first, if the gantry tilt dif-
i
. 2
B
fers from 0°.
B . 2
With the STOP keys, you can interrupt unit movements in an
emergency (table movements and gantry tilting) and switch off
radiation.
If you press one of the red STOP keys, unit movements are inter-
B . 2
rupted and radiation is stopped. The functions of the keys for
system movements are blocked as well. (→ Page A.2-16 Termi-
nating system movements and radiation)
System Description
Emotion Operator Manual