Switching the System On and Off
Restarting the system
D.1-28
syngo CT 2007E
02/2008
Normal power supply is reactivated
D . 1
CAUTION
The detector heating is switched off as long as the system is dis-
connected from the power supply. When you restart the system,
the detector has not yet reached operating temperature!
D . 1
Incorrect diagnosis resulting from image artifacts.
Calibrate the system as part of the checkup. Repeat calibra-
tion after about 60 minutes (Setup > Calibration), or earlier
if ring artifacts occur.
Restart the system. (→ Page D.1-1 Switching the System
On and Off)
Operation
Emotion Operator Manual
D . 1
D . 1
D . 1
D . 1
0 . 0