Read the 300 mm position on the ruler.
Enter the value in the entry field of the content area.
Click on Continue.
F . 3
The table top now moves stepwise into the gantry (300 mm)
and back.
Enter the actual values at the 300 mm and the 0 mm posi-
tion in the entry field again.
Constancy Test
Emotion Operator Manual
F . 3
F . 3
0 . 0
F.3-27
0 . 0
0 . 0