CTDI Test
Phantoms
F.4-2
syngo CT 2007E
02/2008
The measurements are performed with the 16 cm CTDI head
F . 4
phantom and the 32 cm CTDI body phantom. The phantoms are
made of PMMA (perspex). Each phantom has 5 drillings for the
dose chamber. The drillings not occupied by the chamber have
to be filled with the supplied cylinders.
(1) 16 cm CTDI head phantom
(2) 32 cm CTDI body phantom