Quality Assurance
Starting the constancy test
syngo CT 2007E
02/2008
A reference measurement follows the same sequence as a nor-
F . 3
mal constancy test.
Confirm selection of the constancy test with Go.
F . 3
You are prompted to enter the name of the tester and the serial
numbers of the phantoms.
Enter your name and the data required.
Click on Go.
F . 3
Prepare the measurement as described. (→ Page F.1-1
General Information)
Click on Go when you have completed your preparations.
F . 3
The "Quality Assurance Patient" is selected.
You are prompted to do press Start.
Press the Start key on the control box.
The phantom position and the gantry tilt are checked.
F . 3
Constancy Test
Emotion Operator Manual
F . 3
F . 3
F . 3
F . 3
F . 3
F . 3
0 . 0
F.3-7
0 . 0
0 . 0