Constancy Test
Performing the table posi-
tion test
F.3-26
syngo CT 2007E
02/2008
With this quality measurement, you test whether the actual
table position matches the table position displayed.
F . 3
The Table Position test is performed without triggering radia-
tion. It is started with Go in the Quality Constancy dialog box.
(→ Page F.3-2 Performing the constancy test)
After the previous test has been completed, the Table Position
i
. 3
F
test is started automatically.
Mount a ruler at the mobile part of the patient table so that
the 0 mm mark of the ruler is aligned with the stationary
part of the table.
Place the 135 kg weight on the patient table.
Start measurement with Go.
F . 3
The horizontal table position is set to 0. The table top moves
300 mm in the direction of the gantry.
Quality Assurance
Emotion Operator Manual
F . 3
F . 3
F . 3
F . 3
0 . 0
0 . 0