System checkup and calibration

After syngo has started up, you should start system checkup
and calibration to enable the system for scanning.
The checkup procedure is started in the Checkup dialog box
which will be displayed after system startup. If the Checkup dia-
log box does not appear automatically, select Setup > Checkup.
If the last checkup has been completed no longer than 12 hours
i
D . 1
ago, the time since the last complete checkup is indicated.
You have the option of skipping the checkup. If necessary, you
can calibrate the system later (Setup > Calibration).
CAUTION
When you start or restart the system, the detector has not yet
reached operating temperature!
D . 1
Incorrect diagnosis resulting from image artifacts.
Calibrate the system as part of the checkup. Repeat calibra-
tion after about 60 minutes (Setup > Calibration), or earlier
if ring artifacts occur.