Siemens SOMATOM Emotion syngo CT 2007E Operator's Manual page 472

Table of Contents

Advertisement

Constancy Test
F.3-12
syngo CT 2007E
02/2008
Compare the position of the short / long strip on the image
with the options displayed in the content area of the Qual-
ity Constancy dialog box.
Correct the table position by clicking Feed In or Feed Out.
F . 3
With Feed In and Feed Out, you can move the table by 1 mm.
i
F
. 3
Record a new image and compare the position of the strips
again.
Repeat image recording and correction of the table posi-
tion until you have set the correct z-position.
Click on Continue.
F . 3
Quality Assurance
Emotion Operator Manual
F . 3
F . 3
0 . 0
0 . 0