Quality Assurance
Performing the measurement
syngo CT 2007E
02/2008
F . 4
Click on Continue.
F . 4
You are prompted to press Start.
Press the Start key on the control box.
F . 4
Read off the value on the dosimeter and enter it in the pro-
tocol.
Continue with the next kV values.
Then insert the perspex rod with the measuring chamber
into position B of the phantom.
Click on Continue to start the second measurement.
F . 4
The second measurement might not involve all positions tested
i
F
. 4
in the first measurement.
When you have completed CTDI measurement on the CTDI
head phantom continue in the same way with the CTDI
body phantom.
CTDI Test
Emotion Operator Manual
F . 4
F . 4
F . 4
0 . 0
F.4-9
0 . 0
0 . 0