System configuration for calibration
•
Compression paddle must be in
•
Reference dosimeter must be placed under the compression paddle
•
Dosimeter reference plane must be located at 550mm from focal spot
•
Characteristics, response linearization, and unit of measure of reference Dosimeter must be
configured before to start calibration
•
Reference dose outside kV range of reference Dosimeter must be set at ZERO
Radiation table
MAMMOMAT Balance
Troubleshooting Guide
To view and edit the radiation table for the selected mammo
press the "Edit"-"Radiation table" menu.
SPB7-115.840.01
Page 42 of 62
Rev.02
10.04