System Configuration For Calibration; Radiation Table - Siemens MAMMOMAT Balance Troubleshooting Manual

Table of Contents

Advertisement

System configuration for calibration

Compression paddle must be in
Reference dosimeter must be placed under the compression paddle
Dosimeter reference plane must be located at 550mm from focal spot
Characteristics, response linearization, and unit of measure of reference Dosimeter must be
configured before to start calibration
Reference dose outside kV range of reference Dosimeter must be set at ZERO

Radiation table

MAMMOMAT Balance
Troubleshooting Guide
To view and edit the radiation table for the selected mammo
press the "Edit"-"Radiation table" menu.
SPB7-115.840.01
Page 42 of 62
Rev.02
10.04
Dose calculator
Siemens AG
Medical Solutions