OPERATION (CONTINUED)
FIELD CALIBRATION MODE (CONTINUED)
Field Calibration Method
At the beginning of the calibration
method entry menu, the software will
allow the user to start the calibration
process or to exit back to normal
operation. (See Figures 11-1 & 11-2)
Figure 11-1
Figure 11-2
Display Dispense Method
Dispense/Display is a field calibration
method by which the user will dispense
a known volume of fluid and the
software will keep track of the pulse
count during the dispense operation.
Figure 12-1
18
The bottom row of characters will be
in focus to indicate that the user can
select between either "Start" or "Exit" by
advancing focus.
If "Start" is in focus when advancing to the next
menu, the software will automatically advance
to the calibration menu for the calibration
method the user previously selected.
If "Exit" is in focus when advancing to the
next menu, the display will return to normal
operation.
The user will then enter the known
volume into the display and the
software will calculate a K-factor for the
volume dispensed.
The first screen in this calibration method will
show "run 1" indicating that computer is waiting
for flow to start. (See Figure 12-1)
NOTE: Pressing any single button while on this
screen will not have any effect. Pressing both
buttons and holding will allow the user to exit
calibration mode.