OPERATION (CONTINUED)
FIELD CALIBRATION MODE (CONTINUED)
Figure 12-2
When the user is finished with the
run, press and hold the Display button
until the left most digit begins to flash
indicating it is in focus, then release the
Display button.
Pressing the Calibrate button will
increment the digit in focus. Pressing
the Display button will advance focus to
the next digit.
Exit Dispense Display Method
This menu is used to indicate to
the user they are exiting the field
calibration mode menu. The user can
exit any of the calibration methods
at any time. To exit, press and hold
the Display and Calibrate buttons
Figure 14-1
Figure 14-2
Begin dispensing into a container of known
accurate volume. As soon as pulses are
detected by the software, the screen will switch
to display the volume being dispensed on the
top row and the volume unit on the bottom row.
(See Figure 12-2)
Holding Display and pressing Calibrate
will lock in the entered volume. (See
Figure 14-2)
The software will then calculate the
K-factor for the volume entered based
on the pulse count for the run and
the volume entered. The average
frequency and the K-factor will be
entered into the custom user K-factor
table for run 1.
simultaneously for 3 seconds; the
software will display a three second
count down (see Figure 14-1), then
reset to show the exit display. (See
Figure 14-2)
NOTE: After 3 seconds, "FldCAL" will then
be displayed on the top row and "Exit" on the
bottom row (see Figure 14-2) until the user
releases both buttons. The software will store
the sorted table of frequencies and K-factors.