B.2 Receiver - GE Mini Telemetry Operation And Maintenance Manual

Table of Contents

Advertisement

Appendix B

B.2 Receiver

Category
Power Requirements
Nominal Line Voltage:
Line Frequency:
Power Consumption (maximum):
Touch current:
Physical Characteristics
Height:
Width:
Depth:
Weight:
Environmental Conditions
Ambient Temperature:
Relative Humidity:
Atmospheric Pressure:
Altitude:
Certification and Compliance
UL:
FCC:
Industry Canada:
EN:
Output Signals:
48
Technical Specifications
100-120VAC
Fuse 0.25A T
50/60Hz
30 W
<300µA
4" (10.2cm)
8.1" (20.6cm)
11.9" ( 30.2cm)
5.6 lbs (2.55kg)
Storage and Transport
14°F to 131°F (–10°C to 55°C)
5% to 95%, non-condensing
700–1060 mbar (525–795
mmHg)
UL 60601-1
United States FCC 47 CFR part 95
Canadian RSS-210
European EN 300-220-1
US, FECG, MECG, TOCO, IUPC and Mark
2049809-001
Rev B
Receiver
220-240VAC
Fuse 0.25A T
© 2010 by General Electric Company. All rights reserved.
Operating
50°F to 104°F (10°C to 40°C)
5% to 95%, non-condensing
700–1060 mbar (525–795 mmHg)
Up to 3000m