Antenna system

Receiver system

Server

2001989-200E
There are three transmitter configurations that define the ECG lead analysis option
and determine whether other devices, such as oximeters, can be connected to the
transmitter's interface connector ports.
Single-Lead with inactive interface connector ports (blue dust covers).
Single-Lead with active interface connector ports (gray dust covers).
Multi-Lead with active interface connector ports (gray dust covers).
For more information on ECG lead analysis options, refer to
16.
The two INTFC (interface) connector ports are used for connecting serial interface
devices. The ports are labeled 1 and 2 (on the dust covers).
2 is the inside port, closest to the leadwire set. It is for use with continuous
monitoring serial devices, such as oximeters.
1 is the outside port, furthest from the leadwire set. It is for use with episodic
monitoring serial devices, such as blood pressure monitors.
For more information on connecting devices to the transmitter, refer to
components on page
3-7.
The antenna system sends data to the receiver system. Data is then transmitted via a
dedicated Ethernet connection to the server for further processing and viewing. For
more information, refer to the ApexPro CH Telemetry System Antenna Equipment
Reference manual.
The receiver system sends data via a dedicated Ethernet connection to the server for
further processing and viewing. For more information, refer to ApexPro Receiver
System Service Manual.
A server hosts the ApexPro software. The server can be a telemetry server or a CIC
Pro center.
If your system includes a telemetry server, it receives and analyzes patient and
transmitter data from the receiver system, runs the ApexPro software, stores the data
and sends the data to the CIC Pro center for display. For more information, refer to the
appropriate telemetry server service manual.
If your system includes a CIC Pro center, it receives and analyzes patient and
transmitter data from the receiver system, runs the ApexPro software, stores and
displays all telemetry patient data. For more information, refer to the appropriate CIC
Pro center service manual.
ApexPro™
Equipment overview
Lead analysis on page 7-
Optional