GE Mini Telemetry Operation And Maintenance Manual page 51

Table of Contents

Advertisement

Problem
Erratic
FHR/MHR/UA
reading.
FHR/MHR or UA values
are not displayed on the
corometrics monitor when
transducers are plugged
into transmitter.
© 2010 by General Electric Company. All rights reserved.
Troubleshooting
Probable Cause
Transducer
not
positioned
on
abdomen.
Transducer
not
connected to transmitter.
In case of erratic UA reading,
check
if
TOCO/IUP
on the receiver matches the
UA transducer connected to
transmitter.
Receiver
interconnection
cable(s) not properly attached.
Receiver
interconnection
cable(s) defective.
Wrong interconnection cable(s)
in use.
Radio frequency interference.
Another transmitter with the
same frequency is in use within
the same facility.
Exceeding transmission range. •
Shielding effect of hospital
structure.
Transmitter front-end circuitry
not working.
Faulty transducer.
Transmitter and/or receiver off. •
Receiver
interconnection
cable(s) not properly attached.
Receiver
interconnection
cable(s) defective.
Wrong interconnection cable(s)
in use.
Faulty transducer.
Solution
properly
Reposition transducer.
mother's
properly
Ensure the transducer is securely
attached to the transmitter.
Ensure TOCO/IUP switch on the receiver
switch
matches the UA transducer connected
to transmitter.
Ensure interconnection cable(s) is firmly
attached to both monitor and receiver.
Replace interconnection cable(s).
Verify interconnection method.
Instruct patient to stay within signal
range and designated areas where
reception is clear.
Discontinue
transmitters. NOTE: Mini Telemetry can
be re-programmed to an alternative
frequency.
Install optional ceiling antenna system.
If problem persists, contact your
Information
Representative.
Contact your Information technologies
service representative.
Ensure all three devices are turned on.
Ensure interconnection cable(s) firmly
attached to both monitor and receiver.
Replace interconnection cable(s).
Verify interconnection method.
Contact your Information Technologies
Service Representative.
2049809-001 Rev B
Chapter 6: Troubleshooting
use
of
one
of
Technologies
Service
the
37