Specifications
Table 1- 1: Oscilloscope general characteristics
Acquisition
Acquisition Modes
Acquisition Rate, typical
Single Sequence
g
q
Inputs
Input Coupling
Input Impedance, DC
Coupled
P2220 Probe
Attenuation
Supported Voltage Probe
Attenuation Factors
Supported Current Probe
Scales
TPS2000 Series Digital Storage Oscilloscopes Service Manual
These specifications apply to all TPS2000 Series oscilloscopes. To verify that an
oscilloscope meets specifications, it must first meet the following conditions:
H The oscilloscope must have been operating continuously for twenty minutes
within the specified operating temperature range. See page 1- -8.
H You must perform the Do Self Cal operation, accessible through the Utility
menu, if the operating temperature changes by more than 5 °C.
H The oscilloscope must be within the factory calibration interval of one year.
Specifications begin in Table 1- -1. All specifications are guaranteed unless noted
"typical." Specifications that are marked with the n symbol are checked in the
chapter Performance Verification.
Sample, Peak Detect, and Average
Up to 180 waveforms per second, per channel (Sample acquisition mode, no measurements)
Acquisition Mode
Acquisition Stops After
Sample, Peak Detect
Single acquisition, all channels simultaneously.
Average
N acquisitions, all channels simultaneously, N is selectable from
4, 16, 64, and 128.
DC, AC, or GND
1 MΩ ±2% in parallel with 20 pF ±3 pF
1X, 10X
1X, 10X, 20X, 50X, 100X, 500X, 1000X
1 mV/A, 10 mV/A, 20 mV/A, 50 mV/A, 100mV/A, 200 mV/A, 1 V/A, 5 V/A