Tektronix TPS2000 Series Service Manual page 88

Digital storage oscilloscopes
Hide thumbs Also See for TPS2000 Series:
Table of Contents

Advertisement

Maintenance
6- - 26
Table 6- - 4: List of error codes (cont.)
Number
ID
12
Diagnostics
13
Diagnostics
14
Diagnostics
255
Software
Clearing the Error Log. To clear the error log after fixing a problem, follow these
steps:
1. Enable the Service menus as described on page 5- -2.
2. Push the UTILITY button and push the Service option button to display the
Service menu.
3. Push the Service Diag. option button to display the Diagnostics menu.
4. Push the Clear Error Log option button to clear the error log of past
failures. This also sets the Power Cycle count to zero.
TPS2000 Series Digital Storage Oscilloscopes Service Manual
Description
CH3 diagnostics failed
CH4 diagnostics failed
ADG421 #2 diagnostics
failed
Software defect
Probable cause
Acquisition board or power
supply failure
Acquisition board or power
supply failure
Acquisition board or power
supply failure
Contact Tektronix

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TPS2000 Series and is the answer not in the manual?

This manual is also suitable for:

Tps 2012Tps2014Tps2024

Table of Contents