Operation
The following can be set on the display:
–
Positive end expiratory pressure PEEP.
–
Sensitivity Trigger.
–
Ventilation time ratio I:E or inspiration time Ti.
–
Plateau time Tplat %, in % of the inspiration time.
–
AutoFlow (optional).
–
Hose type
The selected hose type must match the hose type in
use. Otherwise a correct volume measurement can-
not be guaranteed.
–
Cuvette type (optional)
22
Activating/setting the trigger
1
Press the key Settings
rameter is displayed.
2
Select the line Trigger on the display and then set
and confirm the value with the rotary knob.
Small value = high sensitivity.
The ventilation mode AC is shown on the display.
Successful patient triggering is indicated by an asterisk
(*) on the left side of the curves window.
Deactivating the trigger
1
Set a value less than 1 L/min or greater than
15 L/min (off is displayed instead of a value).
2
Press the rotary knob to confirm.
The last effective trigger value is adopted by the ventila-
tor when changing from VC-AC to PC-BIPAP or
SpnCPAP.
NOTE
When the Oxylog 3000 plus is in the ventilation mode
VC-CMV and the trigger is set to a value, the ventilati-
on mode changes to VC-AC.
Pocket Guide Oxylog 3000 plus
Oxylog 3000 plus
A
(A) until the trigger pa-