Operation
5
Confirm. The zero calibration starts and the line dis-
plays Busy.
Note the possible warm-up time. During zero calib-
ration, ventilation settings can be changed. After a
successful zero calibration, the line briefly displays
Pass.
6
Attach the CO
sensor (A) back to the cuvette (B).
2
If zero calibration was not successful:
The Oxylog 3000 plus displays the alarm
!!! CO
Zero calib. failed.
2
Repeat zero calibration
If zero calibration is still not possible:
1
Check whether the sensor (A) is soiled and clean if
necessary. If the sensor is defective, replace the
sensor.
2
Repeat zero calibration.
CO
filter check during ventilation
2
A
B
To perform CO
filter check:
2
1
Remove the CO
sensor from the cuvette.
2
2
Attach the CO
sensor (A) to the test filter (B).
2
3
Press the Settings key.
4
Select and activate the line CO2 Filter check - Run.
5
Confirm. The filter check starts and the line displays
Busy.
During the filter check, ventilation settings can be
changed.
After a successful filter check, the line briefly dis-
plays Pass.
6
Attach the CO
sensor (A) back to the cuvette.
2
30
If the check was not successful:
The Oxylog 3000 plus displays the alarm !!! CO2 Filter
check failed. The test value is outside the permissible
tolerance.
1
Check whether the sensor (A) or test filter (B) is
soiled and clean them if necessary. If the sensor is
defective, replace the sensor.
2
Check the CO
calibration with test gas or calibrate
2
the CO
sensor.
2
Pocket Guide Oxylog 3000 plus