Dräger Oxylog 3000 plus Pocket Manual page 24

Software version 1.n
Hide thumbs Also See for Oxylog 3000 plus:
Table of Contents

Advertisement

Operation
Insp.term.%PIF
Inspiration termination criterion of pressure suppor-
ted strokes, as percentage of the peak inspiratory
flow (PIF).
Hose type
The selected hose type must match the hose type in
use. Otherwise a correct volume measurement can-
not be guaranteed.
Cuvette type (optional)
Setting pressure support VC-SIMV/PS
The following can also be set on the display for
VC-SIMV:
Setting on page 1: Pressure support
PEEP.
Setting on page 2: When
0 mbar, the pressure rise time Slope can be set.
Flat slope = long pressure rise time
Medium slope = medium pressure rise time
Steep slope = short pressure rise time.
24
Δ
Psupp above
Δ
Psupp is set above
Setting AutoFlow (optional)
The following can also be set on the display for
VC-SIMV and VC-SIMV/PS:
The AutoFlow function AutoFlow-AF.
When AutoFlow is switched on, the setting Tplat% is no
longer valid, and Slope must be set.
Pocket Guide Oxylog 3000 plus

Advertisement

Table of Contents
loading

Table of Contents